Title - Tender Specification for the - Differential Scanning Calorimeter (DSC) for Food Analysis
Tender Reference Number: TDU/Tender/2026-2027/51
Deadline for Submission: 8th November 2026 at 04:00 PM
Name & Technical Specification:
TECHNICAL SPECIFICATION-Differential Scanning Calorimeter (DSC) for Food Analysis
Purpose and Scope
This document specifies the technical requirements for a heat-flux differential scanning calorimeter (DSC) to be installed in the food analysis laboratory at TDU. The instrument is intended to characterise the thermal behaviour of three principal food components — proteins, starches, and fats/oils — with the following core objectives:
Quantify the degree of protein denaturation (denaturation temperature and residual denaturation enthalpy).
Characterise starch gelatinisation (onset, peak and end temperatures, and gelatinisation enthalpy), and related retrogradation and amylose–lipid complex transitions.
Study fats and oils: melting and crystallisation profiles, solid-fat content, polymorphic behaviour, and oxidative stability (oxidation onset temperature / induction time).
Key Technical Considerations Before Purchase
The following factors should be assessed and confirmed by every bidder. They are the points on which food DSC performance most often succeeds or fails.
Measuring principle — heat flux vs power compensation. Heat-flux (single furnace) instruments, in which sample and reference sit on separate sensors within one furnace, are the mainstream choice and comfortably cover food applications. Power-compensation (double-furnace) instruments give the fastest response and best separation of close-lying peaks, which can help with fat polymorphism, but at higher cost. For this laboratory, heat-flux is sufficient and cost-effective; bidders should state their principle clearly.
Temperature range and cooling. Oils crystallise between −60 and −80 °C, so a cooling facility reaching at least −80 °C without liquid nitrogen is the practical minimum. Liquid-nitrogen cooling (to ≈−150 °C) is a useful option; the upper limit should comfortably be 400 °C for oxidation-onset work.
Sensitivity and resolution. Starch and protein transitions are weak (a few J/g) and fat melting produces multiple overlapping peaks. The instrument therefore needs high sensitivity (0.1 µW) to detect weak events and high resolution to separate close-lying ones. Both must be quoted with the basis of measurement.
Baseline stability and flatness. A flat, repeatable baseline is essential to integrate weak endotherms accurately and to detect small glass transitions. Look for baseline repeatability and baseline stability/curvature figures over the relevant range.
Temperature accuracy and precision. Determines the reliability of onset/peak temperatures. Precision (repeatability) is as important as accuracy for QC use; both should be quoted from repeated indium runs.
Calorimetric (enthalpy) accuracy and precision. Directly governs the accuracy of ΔH, and therefore of the calculated degree of denaturation and gelatinisation enthalpy. Insist on values derived from ten consecutive indium runs.
Heating and cooling rates. A wide, programmable range (e.g. 0.01–100 °C/min or better) supports kinetic studies and method transfer. High cooling rates improve throughput.
Sensor and furnace construction. A multi-thermocouple thermopile sensor (typically 50–120 thermocouples) with a chemically resistant surface (ceramic or gold-coated) is preferred. A corrosion-resistant furnace (silver/silver-alloy or platinum-alloy) ensures durability and cleanliness, especially under oxygen.
Gas control. Integrated mass-flow controllers for method and protective gases, plus software-controlled automatic gas switching between inert (N₂) and reactive (O₂) atmospheres, are required for oxidative-stability testing. An eco/idle mode to shut off gas when idle reduces running cost.
Crucibles and sample sealing. The crucible range matters as much as the instrument. The purchase must include a hermetic sealing press, aluminium open and hermetic pans, large-volume sealed/high-pressure crucibles for starch slurries, pre-pierced lids for oxidation studies, and a sapphire disc for heat-capacity calibration.
Automation and upgradability. The system should be modular and upgradable to an autosampler (typically 34–70 positions) later, so that routine throughput can be increased without replacing the instrument.
Software and data analysis. Software must run on a current Windows platform and evaluate glass transition, melting, crystallisation, onset/peak/end, enthalpy, specific heat capacity, and purity, with modulated (temperature-modulated) DSC to separate reversing from non-reversing heat flow. Quality-control features (control charts) and customisable PDF reporting are desirable; software upgrades should be provided at no extra cost.
Standards compliance. The instrument and methods should align with relevant standards — for example ASTM D3418, E793, E794, E928, E1269, D3895, D4591 and D6604, and ISO 11357 and 19935 — and, where applicable, pharmacopoeial thermal-analysis methods.
Service, training and warranty. On-site installation and application training, an authorised Indian service representative, a clear warranty period, and an option for an annual maintenance contract should all be confirmed.
Infrastructure and operating environment. Pre-installation requirements (power, gases, bench space, ambient temperature), supply of a suitable PC and printer, and any requirement for a cooling unit or liquid-nitrogen dewar should be stated by the bidder.
Minimum Technical Specification
The following table defines the minimum acceptable performance. Bidders must state their actual values and the basis of measurement. Values shown as “better than” indicate a preference for higher performance where it is available at reasonable cost.
Parameter | Minimum acceptable requirement | Rationale / notes |
Type | Heat-flux or Power-compensation DSC; sample and reference on separate sensors within one furnace | Mainstream, robust design covering all three food applications |
DSC Furnace Design | It should be Silver Alloy Furnace or Pt/Ir furnace or better | For high performance over longer life. |
Temperature range (standard config.) | - 80C to 500C or better using Intracooler or ECU or refridgeration unit. | Covers oil crystallisation (−60/−80 °C) through to oxidation onset (>400 °C) |
Temperature range with Intercooler or chiller or cooling unit | Upgradable to −150 °C (liquid nitrogen) and 600–700 °C. Preferably DSC shall have flexibility to connect both the attachment simultaneously depending on application in future | Future-proofing; low-temperature and high-temperature extensions |
Cooling | Built-in electrical/mechanical cooling to at least −30 °C without liquid nitrogen | Avoids dependence on LN₂ for routine oil work |
Temperature Accuracy | +/- 0.05 to 0.1 C |
|
Temperature Precision | Better than 0.02 C |
|
Heating rate | 0.01 to 100 °C/min (up to 300 °C/min desirable) | Method flexibility and kinetic studies |
Cooling rate | ≥ 50 °C/min (up to 150 °C/min desirable) | Throughput and controlled cooling for crystallisation |
Sensor | ≥ 50 thermocouples (thermopile), It should be Chemical and Corrosive Gas resistant. It should be either Gold coated Thermopile or Pt/Ir or Ni/Cr alloy with longer life. | High sensitivity and resolution for weak starch/protein transitions |
Furnace material | Corrosion-resistant silver / silver-alloy or platinum-alloy construction | Durability and cleanliness, especially under oxygen |
DSC sensitivity | 0.1 µW | Detects weak gelatinisation and denaturation events |
Digital resolution | 0.02 to 0.05 µW | Fine resolution of small heat-flow changes |
DSC Noise | +/- 0.1 to 0.2 µW |
|
Measurement / dynamic range | ± 170 mW to 500 mW to cater to all types of samples | Handles large melting/oxidation events without saturation |
Baseline repeatability | ± 5 µW or lower | Reliable integration of weak endotherms |
Baseline curvature/stability / flatness | < 20 µW over −50 to +300 °C | Accurate baselines across the food working range |
Temperature accuracy | ≤ ± 0.1 K (better ± 0.05 K), from indium runs | Reliable onset/peak temperatures |
Temperature precision | ≤ ± 0.01 K | Repeatability for QC and comparability studies |
Calorimetric (enthalpy) accuracy | ± 0.5 % or better, from 10 consecutive indium runs | Accuracy of ΔH → degree of denaturation and gelatinisation enthalpy |
Calorimetric precision | ± 0.05 % or better | Reproducibility of enthalpy measurements |
Indium response ratio | > 12 mW/°C (1 mg In at 10 °C/min) | Standard sensitivity benchmark; higher is better |
Gas control – Mass Flow Control for two gases | It should be supplied with built in two gas mass flow controllers and it should allow change of gas during the run and method getting stored with data file. Software-controlled automatic switching between N₂ and O₂; eco/idle mode | Essential for oxidative-stability (OOT/OIT) testing; reduces gas cost |
Gas cylinders with regulators | Suitable gas cylinders of N2 and O2/Air gases of IOLAR grade II quality with dual stage gas regulator, Moisture traps shall be quoted by supplier |
|
Crucibles / sample prep; Universal press to crimp all types of pans. | Universal/Hermetic sealing press; Al crucibles 400 No’s with lids and 200 hermetically sealed crucibles for 70 microlit capacity; Al pans/covers of 5MPa pressure; large-volume sealed/high-pressure crucibles (40–70 µL); pre-pierced lids; sapphire disc for Cp. | Retains water for starch/protein work; enables oxidation and Cp studies |
Calibration standards | ≥ 3 traceable reference materials (e.g. In, Sn, Zn) with certificates, plus sapphire | Accurate, traceable calibration |
Automation | Provision to upgrade to an autosampler (≥ 34–50 positions) later | Future throughput without replacing the instrument |
Software | Windows 11/10-compatible or better; Evaluation of phase transition, Analysis software should have provision to do evaluations such as Tg, melting, crystallisation, onset Delta Y/peak & peak area/end, peak temperature, overlay, insert signal, slope correction, display of all temp segments, baseline correction, subtraction, calibration routines, DDSC graphs, isocrystallization studies, OIT test, enthalpy, specific heat Cp (single run), crystallinity, purity, kinetics, modulated DSC, QC/control charts, customisable PDF reports; free version upgrades.
This software shall be also compatible with TGA/STA,TMA and DMA for future connections and operations. Original licensed version of the software to be provided in electronic media format like CD or dongle
Supplier shall quote various advanced software like Sine wave Modulated DSC and this shall be part of the package and will be supplied with DSC unit | Complete evaluation of all three application families |
Standards | ASTM D3418, E793, E794, E928, E1269, D3895, D4591, D6604; ISO 11357, 19935; | Method credibility and traceability |
Display | Built-in status display (eco / ready / measuring / warning / error) preferred | Convenient standalone operation |
Warranty | Minimum 3 years comprehensive (from date of installation) | Protects the investment; standard in academic tenders |
Training | On-site installation and application training by a qualified engineer, at no extra cost | Ensures competent, independent use |
PC & printer | Intel Core i7, 16 GB RAM, 1 TB SSD, Windows 11/10, MS Office, Adobe Acrobat, 23-inch monitor, keyboard, mouse; laser MFP printer | Complete, ready-to-use workstation |
Upgrades | Photocalorimeter upgrade and camera attachment upgrade should be available with vendor |
|
Accessories and Consumables to be Quoted
The following must be included in the offer (not quoted as unbudgeted extras):
Hermetic crucible sealing press and sealing tools.
Crucible handling / accessory set (tweezers, spatula, funnel, needles, cutter and cutting plate, micro-syringe, glass-fibre pen, air blower, push-rod).
Aluminium open pans and crimp covers; aluminium hermetic (sealed) pans.
Large-volume sealed or high-pressure crucibles (40–60 µL) for starch and protein slurries.
Pre-pierced / pierceable lids for oxidation and autosampler use.
Sapphire disc (and/or the manufacturer's Cp reference) for specific-heat calibration.
Calibration standards with traceability certificates (at least three, e.g. indium, tin, zinc).
Gas tubing, fittings and pressure regulators for purge/protective gases.
Workstation PC and laser multifunction printer (as specified in Section 4).
Recommended: analytical balance (0.01 mg readability) and UPS for the instrument and PC.
General, Commercial and Service Terms must be included
Pre-installation requirements to be stated by the bidder (power supply, gases, bench space, ambient conditions, cooling-water or LN₂ provision).
Instrument manual to be supplied in both hard copy and electronic form; a detailed standard operating procedure (SOP) for routine analysis to be provided.
Authorised Indian representative: the bidder must name its authorised Indian representative with contact details, and provide a list of existing users of similar DSC/thermal-analysis installations in India with contact details.
Installation and on-site training by a qualified installation/application engineer, covering routine analysis, method development, calibration and routine maintenance, at no extra cost.
Warranty of at least three years, comprehensive, from the date of installation.
Software updates/upgrades to be available at no additional cost.
Delivery, payment terms, cancellation and service/AMC terms to be clearly stated.
Compliance with applicable local procurement norms and after-sales support availability to be confirmed.
Tender Submission:
The tender will follow a two-bid system, consisting of:
1) Technical Bid: Include all technical details, specifications, and relevant documentation.
2) Financial Bid: Submit detailed pricing information.
Submission Instructions:
1) Bidders must submit their bids in separate sealed envelopes clearly marked as "Technical Bid" and "Financial Bid."
2) Both envelopes should be placed in a larger sealed envelope clearly mentioning : Tender Title & Tender Reference Number
Submission Deadline:
1. All bids must be submitted by 8th November 2026 at 04:00 PM
2. Late submissions will not be considered.
Submissions can be made in person/courier at:
TDU
No. 74/2, Jarakabande Kaval, Post Attur
Yelahanka, Bengaluru 560064
Tel: 080 2856 8000
Additional Documents to be included under Technical Bid:
1) Scanned copy of Firm's Registration under Shops and Establishment Act
2) PAN Card Copy
3) GST Registration Certificate
4) Aadhaar Card Copy (incase of individual)
5) Bank Details of the Firm
6) Last 3 years Audit Statement of the company
7) Latest 3 contact details of the customers who purchased the product from the vendor and with their review on the products.
Contact Information:
For further information and clarification, please contact:
Mr. Naresh N K · nareshnk@tdu.edu.in · +91 9980213805